ASTM F2009-00(2005)
Historical Standard: ASTM F2009-00(2005) Standard Test Method for Determining the Axial Disassembly Force of Taper Connections of Modular Prostheses
SUPERSEDED (see Active link, below)
ASTM F2009
1. Scope
1.1 This test method establishes a standard methodology for determining the force required, under laboratory conditions, to disassemble tapers of implants that are otherwise not intended to release. Some examples are the femoral components of a total or partial hip replacement or shoulder in which the head and base component are secured together by a self-locking taper.
1.2 This test method has been developed primarily for evaluation of metal and ceramic head designs on metal tapers but may have application to other materials and designs.
This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.
2. Referenced Documents (purchase separately) The documents listed below are referenced within the subject standard but are not provided as part of the standard.
ASTM Standards
E4 Practices for Force Verification of Testing Machines
F1636 Specification for Bores and Cones for Modular Femoral Heads
Keywords
arthroplasty; disassembly; heads; hip prosthesis; modular; shoulder prosthesis; tapers; Axial disassembly force; Ceramic head designs; Disassembly; Force; Locking mechanisms (of tapers); Metal head designs; Modular prostheses; Testing methods--surgical implants; Prostheses; Surgical implants;
ICS Code
ICS Number Code 11.040.40 (Implants for surgery, prothetics and orthotics)
DOI: 10.1520/F2009-00R05
ASTM International is a member of CrossRef.